Fluke recall

Friday, 16 April, 2004

In cooperation with the U.S. Consumer Protection Safety Commission, Fluke Corporation is voluntarily recalling the TL221, TL222 and TL224 modular test leads.

Fluke is asking users to send back test leads for replacement. This recall does not affect the test leads permanently attached to test probes that come standard with most Fluke Digital Multimeters. The recall does not include currently manufactured TL221, TL222 and TL224 test leads that are marked with a SureGrip(tm) symbol, either by a tag around the silicon wire, or the symbol molded onto the connector end.

The recall only impacts test leads TL221, TL222, and TL224. They are found in the following accessories:

TL220 TL223 TLK220 TLK225 L215

The affected leads are also included with the following Fluke equipment with serial numbers between 83260001 and 85210001:

Fluke 179/1AC Fluke 179/EDA

Fluke 88 Fluke 741B

Fluke 743B Fluke 744

Description of problem

TL221, TL222 and TL224 test leads, manufactured between December 2002 and October 2003, may have a faulty connection that could cause the tester to not indicate that voltage or current is present. This could place the user in a potentially hazardous situation.

How to identify recalled test leads

There are only three Fluke test lead models affected:

TL221 Straight to Straight

TL222 Right to Right

TL224 Right to Straight

They have the following characteristics:

The leads have only been available for purchase since January 2003.

These are modular leads with no probes attached to the end.

The connectors have a black overmolding.

The safety rating inscribed near the logo is CAT III 1000 V, CAT IV 600 V.

The Fluke logo can be found on the connector ends.

Actions

Stop using the affected TL221, TL222 and TL224 test leads.

Verify that your leads are affected by this recall by visiting our website.

Return your test leads to your local Fluke sales representative for free replacement.

Please do not send your DMM, other accessories, or non-affected test leads.

For more information about the recall visit Fluke's website: www.fluke.com/TL22Xreturns.

Contact

For more information about the Fluke Corporation, or to find the location of the nearest distributor, visit our Web site at www.fluke.com.au e-mail sales@fluke.com.au or call 02) 8850 3333, fax 02) 8850 3300.

Item provided courtesy of Fluke Corporation

Related News

UV broadband spectrometer enhances air pollutant analysis

The laser-based technology developed at TU Graz enables the continual real-time analysis of air...

Physicists produce signals with attosecond duration

A team of physicists has used paired laser pulses to compress a series of electron pulses to a...

New microchip developed using two Nobel Prize-winning techniques

Researchers have developed on-chip technology to measure distances in materials at high...


  • All content Copyright © 2024 Westwick-Farrow Pty Ltd