Current switching and test system accuracy

Scientific Devices Australia
Wednesday, 07 July, 2010


The differences among the types of signals that a test system’s switching hardware must handle are not always well understood. Unless these differences are taken into account in switch system design, they can degrade signal integrity and affect overall test system performance. This article reviews switching techniques for maximising the integrity of current signals.

Voltage sources are usually able to provide a compliance current up to the programmed voltage. As a result, the typical default condition of a voltage switch is open (in other words, drawing very little current or having high impedance).

Current switching, however, usually requires the default configuration to be a complete circuit. This means the current needs a complete path until switched.

Typically, the switching component (ie, relay) is a normally closed relay or the HI and LO terminals are shorted in the default condition.

Most current measurement applications demand that all current paths be continuous, even when a particular current signal is not connected to the ammeter.

To accomplish this, switch cards designed for current switching often use SPDT or form C relays (see Figure 1).

 
Figure 1: Using form C relays to switch multiple current sources.

As illustrated in Figure 1, when a channel is in an open state (ie, not connected to the output), signal HI is connected to signal LO through the normally closed (NC) contact.

In this case, only signal HI is switched. Signal LO is hard-wired to the ammeter and to all other signal LOs.

This is typically the case when testing components with relatively low currents (eg, 1 µA). When a channel is selected, the current source is connected to the ammeter through the normally open (NO) contact.

Note that the current will be interrupted briefly when the form C relay is actuated. This could cause problems when used with high-speed logic or other circuits sensitive to a momentary break in the current flow.

Such a problem can be overcome by using a switch card with a pair of form A isolated switches to provide a make-before-break connection, as illustrated in Figure 2.

 
Figure 2: Using form A relays for uninterrupted current switching.

The 3740 32-channel isolated switch card, designed for use with the Series 3700 switch system/multimeter combines the best of both worlds, offering 28 general-purpose form C channels that are suitable for routing power; for higher power applications of up to 7 A, the card includes four additional high-current form A channels.

A variety of switching topologies is suitable for use in current switching applications:

  • Scanner switching - The scan configuration or scanner is the simplest arrangement of relays in a switch system. It can be thought of as a multiple position selector switch;
  • Multiplex switching - Like the scan configuration, multiplex switching can be used to connect one instrument to multiple devices (I:N) or to connect multiple instruments to a single device (N:1). However, the multiplex configuration is much more flexible than the scan configuration. Unlike the scan configuration, multiplex switching enables making multiple simultaneous connections and also permits either sequential or non-sequential switch closures;
  • Matrix switching - The matrix switch configuration is the most versatile because it can connect multiple inputs to multiple outputs. A matrix is useful when connections must be made between several signal sources and a multi-pin device, such as an integrated circuit or a resistor network.

When designing a switching circuit for high current (>1 A), pay particular attention to the maximum current, maximum voltage and VA specifications of the switch cards and relays.

Also, it is important to choose a switch card or relay with low contact resistance to avoid excessive heating, which can cause contacts to weld together and lead to contact failure. Contact heating is caused by I2R power dissipation.

High-current switching can be used for either switching a power supply to multiple loads or for switching an ammeter to multiple sources.

Figure 3 illustrates switching a power supply to multiple loads using a multiplexer scanner card. In this scenario, the power supply will output 1 A to each of four loads. This does not present a problem when only one channel is closed at a time.

 
Figure 3: Switching a power supply to multiple loads.

However, when all four channels are closed, the power supply will output 4 A through the common path. Unfortunately, even though the maximum current of a particular channel is specified at 1 A, the common path on the switch card may not be able to tolerate 4 A.

This is not usually specified for a switch card, but the limitation is usually a function of the trace width and connector ratings. One way to avoid this problem is to use a switch card with independent (isolated) relays and to make connections with wires rated to carry the total current.

Many electrical test systems are capable of producing hazardous levels of power. These high power levels make operator protection a priority. Some protection methods include:

  • Designing test fixtures that prevent operator contact with hazardous circuits;
  • Double-insulating all electrical connections that an operator could touch;
  • Using high-reliability fail-safe interlock switches that disconnect power sources when a test fixure is opened;
  • Providing proper training to all users so that they can understand potential hazards and know how to protect themselves from injury.

The term ‘cold switching’ indicates that a switch is activated with no signal applied. Therefore, no current will flow when the switch is closed and no current will be interrupted when the switch is opened.

When hot switching, voltage is present and current will flow the instant the contacts close. When the switch is opened, this current will be interrupted and can cause arcing.

Cold switching allows power to be applied to the device under test in a controlled manner. Its primary advantage is longer switch life expectancy than with hot switching. Cold switching also eliminates arcing at the relay contacts and any RFI that might be caused by arcing.

Hot switching might be necessary if close control must be exercised in the period between the application of power and the making of the measurement. For example, hot switching is typically used where digital logic is involved, because devices might change state if power is interrupted for even a moment.

With relatively large relays, hot switching might also be necessary to ensure good contact closure. The connection might not be reliable without the ‘wetting’ action produced by the current flowing through the contacts.

When switching currents of 1 µA or lower, special techniques must be used to minimise sources of interference such as offset currents, leakage currents, electrostatic interference, triboelectric currents and electromechanical currents. Such interference might derive from the switch card itself, the connecting cables or the test fixturing.

Offset currents - offset currents are spurious currents generated by a switching card even though no signals are applied. They are usually caused by galvanic sources on the switch card. Offset current is especially significant when measuring low currents if the magnitude of the offset is comparable to that of the low current being measured.

Leakage currents - leakage current is an error current that flows through insulators when a voltage is applied. It can be found on the switch card, in cabling and in test fixtures.

Even high-resistance paths between low-current conductors and nearby voltage sources can generate significant leakage currents.

To reduce these effects, always use a switch card with high channel isolation and use the guard capability of the measurement instrument.

Another method to help reduce leakage current is to keep the switch card clean. Dirt, body oils and the like will create a low-resistance path and allow leakage currents to flow.

To reduce these currents in the text fixturing, always use good-quality insulators such as Teflon and polyethylene; avoid materials such as nylon and phenolics, which can absorb moisture, affecting their insulating performance.

Electrostatic interference - shielding is required because high-impedance circuitry is susceptible to picking up spurious radiated noise. Relay contacts should be shielded from the coil to minimise induced noise from the relay power supply.

The device under test and interconnect cabling should also be shielded to prevent noise pickup. All shields should be connected to circuit LO.

Triboelectric currents - triboelectric currents are generated by charges created by friction between a conductor and an insulator, such as between the conductor and the insulation of a coaxial cable. The friction can be reduced by using special low-noise cables with conductive coating (such as graphite) and by securing the interconnect cabling to minimise movement.

Electrochemical currents - electrochemical currents are generated by galvanic battery action that results from contamination and humidity.

Cleaning joints and surfaces thoroughly to remove electrolytic residues (which include PC etchants, body salts and processing chemicals) will minimise these effects.

When a relay opens or closes, a charge transfer of the order of pico-coulombs occurs, which causes a current pulse in the circuit. This charge transfer is due to the mechanical release or closure of the contacts, the contact-to-coil capacitance and the stray capacitance between signal and relay drive lines.

After a relay is closed, it is important to allow sufficient settling time before taking a measurement. This can be as long as several seconds, depending on the relay used.

If a step voltage is applied to the circuit, a transient current is generated. This will gradually decay to a steady-state value.

The time needed to reach the steady value (or ‘settling time’) can be used to determine the proper delay time for the measurement.

When designing a measurement system, selection of the switch is as critical as the selection of system instrumentation or the design of the test interface.

The intended application must be thoroughly considered and the switch selected must meet the requirements of the application.

Careful attention to detail and to the basic principles of measurement can help ensure greater system accuracy and performance.

By Dale Cigoy, senior applications engineer, Keithley Instruments

Related Articles

Smart buoy tech live streams data 24/7 from sea floor

A Perth-based company has developed the Nodestream protocol, a bespoke solution that uses...

Engineers discover a new way to control atomic nuclei as “qubits”

Using lasers, researchers can directly control a property of nuclei called spin, that can encode...

Simultaneously locating multiple defects on microchips

Researchers have developed a method that can simultaneously locate individual electrical flaws in...


  • All content Copyright © 2024 Westwick-Farrow Pty Ltd